Heath-consultants Detecto-Pak 4 (DP) Uživatelský manuál Strana 17

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 49
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 16
100786-0 Rev. D 05/09
10
CHAPTER THREE
III. CALIBRATION MOTHER TANKS/CALIBRATION
CYLINDERS:
This Chapter discusses, in detail, the proper steps to take in transferring
calibration gas to a calibration cylinder.
Recommendations for ordering calibration gas mother tanks, along with
proper storage, are also discussed.
A. CALIBRATION GAS MIXTURES
The calibration gas is comprised of a mixture of 100 ppm + 10% methane,
balance zero air. The calibration gas mixture is drawn into the instrument
at the inlet of the lter housing to ensure proper sensitivity level (See
Chapter Six for a detailed explanation of this method.)
B. TRANSFILLING CALIBRATION CYLINDERS
Heath assumes no liability in the relling of calibration cylinders. If you
wish to rell your own calibration cylinders, caution must be exercised
during the relling process. It is also recommended that all safety precau-
tions be used in the handling of high-pressure compressed, ammable
gases during the relling process.
To rell the calibration cylinder, connect the calibration transller to
a calibration mother tank, then attach a non contaminated calibration
cylinder to the Transller. See Figure 3 this Section and **NOTE**,
Section II, C of Chapter Two.
Zobrazit stránku 16
1 2 ... 12 13 14 15 16 17 18 19 20 21 22 ... 48 49

Komentáře k této Příručce

Žádné komentáře