102105-0 DP-IR MANUAL REV M
Page 24 of 35
FLOW ERROR icon is on
Very slow to purge out
• Filter needs to be changed
• Plugged probe
• Internal leak
• Check lters and replace if
necessary
• Clean probe lter and clear or
replace the tubing
• Check tightness of internal
Hydrophobic lter
Maintenance
In order to maintain the DP-IR in good working condition, the following maintenance should be
performed as indicated:
Maintenance Item Frequency
Self-Test and Calibration Daily to ensure the instrument is
functioning properly
Battery recharge Recharge to full capacity after each use
Replace dust lter Daily or more often as needed depending on
conditions
**Note**
There is a hydrophobic lter inside the front cap behind the dust lter. Replace this lter should
it clog due to water or dust. There is a back up hydrophobic lter inside the instrument housing
to prevent water damage to the instrument. The exterior lters should be changed as needed by
the user.
**Note**
Do not operate the instrument without the dust lter or without the hydrophobic lter.
**Warning**
To reduce the risk of ignition of a ammable or explosive atmosphere, batteries
must be changed or recharged only in a location known to be non-hazardous.
**Caution**
Only use HEATH battery pack P/N 102961-0.