6-11
BARHOLE TESTING MODE
To switch ON the flashlight:
Press and hold the Centre button
again.
The backlight and flashlight are both timed to switch OFF
after one minute. Both can be switched OFF by a third
press and hold the Centre button
.
6.6.3 Flow Fault
If a flow fault is detected, the pump stops automatically.
If a sample was in progress then current sample is halted
and Purge cycle will be the next part of the sequence.
If flow fault is indicated, the instrument should be checked
for water ingress or blockage.
To clear the flow fault:
(Once the blockage has been cleared)
Press the RH button .
6.6.4 Bead Fault
If a bead fault is detected, the pump stops automatically.
If a sample was in progress then current sample is halted
and Purge cycle will be the next part of the sequence.
A bead fault indicates a faulty sensor and therefore the
instrument should be returned for service / repair.