Heath-consultants Remote Methane Leak Detector Intrinsically Safe (R Uživatelský manuál Strana 24

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 40
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 23
When the Self-Test fails. The following table may help to isolate the problem:
Cause
Corrective Action
Low signal strength or noisy
Make sure battery is fully charged
Make sure that the unit is in its proper position, and
all the way down and flat
Initial ppm-m too low or too
high
Make sure battery is fully charged
Make sure the unit is in its proper position, all the
way down, and flat
Verify calibration cell is not broken
Use the Calibration Override
Laser adjustment failed
Make sure battery is fully charged
Make sure that the unit is in its proper position, and
all the way down and flat
Verify calibration cell is not broken
**Caution**
Should the instrument not pass after several attempts, do not use the instrument for
survey work until the problem is corrected. Contact HEATH for further assistance.
101296-0 RMLD IS MANUAL REV J Page 24 of 40
Zobrazit stránku 23
1 2 ... 19 20 21 22 23 24 25 26 27 28 29 ... 39 40

Komentáře k této Příručce

Žádné komentáře