Excessive false detections
while scanning at longer
distances
Scanning too fast
Alarm detection
threshold set too low
Scanning at the range
limit of the instrument
Slow down the
scanning rate. Pause at
the long range and
sweep towards you
Increase the Alarm
detection threshold
Move in closer
Excessive false detections
while scanning at closer
distances
Scanning too fast
Alarm detection
threshold set too low
Avoid making abrupt
motions while scanning
Increase the Alarm
Detection Threshold
Excessive false detections
or loss of sensitivity
Laser output not
optimized
Alarm Detection
Threshold set too high
or low for conditions
Perform Self-Test
procedure
Check the Alarm
Detection Threshold
Error icon or Warning icon
on continuously
Low battery
Moisture condensation
on mirror due to rapid
change in temperature
Internal component
failure
Check battery level and
recharge if necessary
Allow for the
temperature of the
instrument to stabilize
Note error code and
contact HEATH
Battery indicator does not
show full charge after
charging
Battery not fully
charged
Battery level
calibration error
Charge unit until solid
green light on charger
is on
Run unit until it shuts
off, then fully recharge
with out interruption
Battery indicator bars are
flashing on and off
Battery level
calibration error
Run unit until it shuts
off, then fully recharge
with out interruption
101296-0 RMLD IS MANUAL REV J Page 32 of 40